Focused Ion Beam (FIB): Milling, TEM Lamellae and FIB-SEM Imaging

· 8 min read

A focused ion beam (FIB) uses a finely focused stream of gallium or noble gas ions to image a sample and remove material from it with nanometre precision. Combined with an SEM, it cuts site-specific TEM lamellae, exposes clean cross-sections and builds 3D FIB-SEM tomograms. This guide explains how it works, the main workflows and how to limit beam damage.

What a focused ion beam is and why microscopists use it

A focused ion beam (FIB) instrument works much like a scanning electron microscope, except the probe is a beam of heavy ions instead of electrons. Because each ion carries tens of thousands of times more mass than an electron, it does not just generate a signal when it hits the sample: it physically knocks atoms out of the surface. That single property turns the FIB from an imaging tool into a nanoscale machining tool.

In practice the FIB is rarely used alone. Most laboratories run a dual-beam FIB-SEM, where an ion column and an electron column point at the same spot on the sample, usually at an angle of about 52 to 54 degrees to each other. The ion beam cuts, the electron beam watches, and the operator can switch between them without moving the specimen.

If you are new to electron beam instruments, our overview of how the scanning electron microscope works covers detectors and signals that the FIB-SEM borrows directly.

Ion sources: gallium LMIS and plasma FIB

Gallium liquid metal ion source

The classic FIB uses a gallium liquid metal ion source (LMIS). Gallium melts at about 30 °C and wets a sharp tungsten needle. A strong electric field pulls the liquid into a so-called Taylor cone with a tip radius of only a couple of nanometres, and atoms at that tip are ionised and extracted. The result is a very bright, very small source with a narrow energy spread, which is exactly what is needed to focus the beam to a spot of a few nanometres.

Ions are typically accelerated to 30 kV for milling and imaging, while lower energies (5 kV, 2 kV or even below) are used for final polishing. Imaging resolution with gallium is limited to roughly 5 nm, largely because the beam erodes the very features it is imaging.

Plasma FIB (xenon and other gases)

A plasma FIB (PFIB) generates ions from an inductively coupled plasma, most often xenon, though argon, oxygen and nitrogen sources are also offered on some instruments. The plasma source cannot be focused as tightly as a gallium LMIS, but it delivers far higher beam currents, reaching into the microampere range. That makes it the tool of choice when you need to remove large volumes: cross-sections hundreds of micrometres wide, large-volume tomography or packaging failure analysis. Noble gas ions also avoid gallium implantation, which matters for materials such as aluminium alloys where gallium segregates to grain boundaries.

FeatureGallium FIB (LMIS)Plasma FIB (Xe and others)
Best spot sizeA few nmLarger, typically tens of nm at low current
Maximum beam currentTens of nAUp to the µA range
Material removal rateModerate, ideal for precise workVery high, ideal for large volumes
Implanted speciesGa, chemically active in some metalsInert noble gas (for Xe and Ar)
Typical jobsTEM lamellae, nanofabrication, circuit editBig cross-sections, large-volume 3D, cryo work on thick samples

Focused ion beam milling: how material is removed

Focused ion beam milling relies on sputtering. When an ion hits the surface, it sets off a collision cascade in the top few nanometres, and some surface atoms receive enough energy to escape. By scanning the beam over a defined pattern (a rectangle, a line, a circle or a bitmap), the operator removes material only where it is wanted.

Beam current controls the trade-off between speed and precision. A large current removes material quickly but has a wide tail that rounds edges; a small current gives crisp walls but is slow. A typical workflow therefore uses a coarse cut at high current, followed by progressively finer cleaning cuts.

Gas injection: deposition and enhanced etching

Most instruments also carry a gas injection system (GIS). A precursor gas such as a platinum or tungsten organometallic is released close to the surface, and the beam decomposes it locally, writing a metal or carbon deposit exactly where the beam scans. This is used to lay down protective caps before milling, to weld samples to manipulators and, in semiconductor labs, to rewire circuits. Other gases enhance etching of specific materials or reduce redeposition.

Focused ion beam lithography

Because the beam can remove or add material without a mask, focused ion beam lithography is used for rapid prototyping of nanostructures: nanopores, plasmonic antennas, cantilevers and photomask repairs. It is too slow for mass production, but for one-off devices and research prototypes it is hard to beat.

TEM lamella preparation and lift-out, step by step

The most common reason a microscopy facility buys a FIB-SEM is to prepare thin specimens for transmission electron microscopy. A TEM needs a sample thin enough for electrons to pass through, typically below 100 nm, and the FIB can cut that slice from a precisely chosen site: a single grain boundary, a defect, a coating interface or one transistor. Our article on transmission electron microscopy explains what happens to the lamella next.

  1. Locate the site with the electron beam, to avoid ion damage before you start.
  2. Deposit a protective cap (usually platinum or carbon) over the region of interest, first with the electron beam and then thicker with the ion beam.
  3. Mill trenches on both sides of the cap at high current, leaving a wall about 1–2 µm thick.
  4. Undercut the wall by tilting the stage and cutting its bottom and one side free.
  5. Attach the micromanipulator needle with a small deposition weld and cut the last connection to the bulk.
  6. Transfer the chunk to a TEM half-grid post and weld it there, then detach the needle.
  7. Thin both faces with decreasing current until the lamella is electron-transparent.
  8. Polish at low kV (for example 5 kV and then 2 kV) to strip most of the damaged surface layer.

Plan the final thickness with the damage layer in mind: on silicon, 30 kV gallium milling can leave an amorphous layer of roughly 20–30 nm per side, while a 2 kV polish brings it down to a few nanometres. Skipping the low-kV step is the most common reason a lamella that looks perfect in the SEM gives poor high-resolution TEM images.

FIB-SEM tomography: slice and view in 3D

Focused ion beam scanning electron microscopy tomography, often called slice and view, alternates two steps: the ion beam removes a slice a few nanometres to tens of nanometres thick, and the electron beam images the freshly exposed face. Repeating this hundreds or thousands of times builds a stack that software aligns and reconstructs into a 3D volume.

The method is used for pore networks in batteries and catalysts, phase distribution in alloys, and cell and tissue ultrastructure in life science. Besides secondary and backscattered electron images, many systems collect EDS maps at each slice (see our guide to EDS in the SEM), giving 3D chemistry as well as morphology.

  • Strength: nanometre-scale voxels over volumes tens of micrometres across.
  • Limitation: the method is destructive, so the volume is gone once imaged.
  • Typical artefact: curtaining, vertical stripes caused by uneven milling of hard and soft phases.
  • Mitigation: protective caps, rocking-stage milling and destriping algorithms.

Beam damage, gallium implantation and other artefacts

Every ion that removes atoms also disturbs the atoms it leaves behind. The main side effects are:

ArtefactWhat happensHow to reduce it
AmorphisationCrystalline surface turns amorphous to a depth that grows with beam energyLow-kV final polishing; low-energy argon milling afterwards
Ion implantationGallium is embedded in the top nanometres; in some metals it diffuses to grain boundariesUse a plasma FIB with inert ions, or cryo conditions
RedepositionSputtered atoms settle on nearby wallsCleaning cuts, sensible scan strategy, etching gases
CurtainingStripes on cross-sections from uneven sputter ratesThicker protective cap, rocking milling, back-side milling
Heating and chargingBeam-sensitive or insulating samples degrade or driftLower current, conductive coating, charge neutralisation, cryo stage

For biological samples, cryo-FIB mills vitrified cells into thin lamellae under cryogenic conditions, which is now a key step before cryo-electron tomography. Our article on cryo-EM explains why the sample must stay frozen throughout.

A short history of the focused ion beam

The building blocks came together in the 1970s. In 1975 Krohn and Ringo demonstrated a high-brightness liquid metal ion source, and the first FIB based on such a source was reported a few years later. During the 1990s commercial instruments spread through the semiconductor industry for failure analysis and circuit edit, and the combination with an SEM column turned the FIB into a general laboratory tool. Plasma sources and cryo stages are the most important additions of the past decade.

Conclusion: when a FIB is the right tool

Choose a focused ion beam when you need to see or prepare something at a specific location below the surface: a site-specific TEM lamella, a clean cross-section through a coating or device, or a 3D volume of a few tens of micrometres. Choose a gallium source for precision and a plasma source for volume. Whatever the instrument, the quality of the result depends less on the first big cut than on careful low-kV finishing.

Frequently asked questions

How does a focused ion beam work?

An ion source, usually gallium, emits ions that are accelerated to a few to about 30 kV and focused by electrostatic lenses onto the sample. The impacting ions sputter surface atoms and release secondary electrons, so the same beam can both mill and image.

What is the difference between FIB and SEM?

An SEM scans electrons, which image the surface without removing material. A FIB scans much heavier ions, which sputter atoms away, so it is mainly used for milling, deposition and sample preparation. Dual-beam FIB-SEM instruments combine both.

How thin is a FIB-prepared TEM lamella?

For most TEM work the lamella is thinned to below 100 nm, and for high-resolution imaging often to a few tens of nanometres. Final polishing at low kV is needed to remove most of the damaged surface layer.

Why is gallium used in focused ion beams?

Gallium is liquid close to room temperature and has a low vapour pressure, so it forms a stable liquid metal ion source with a very small emission point. That gives a bright beam that can be focused to a few nanometres.

What is a plasma FIB used for?

A plasma FIB, typically using xenon, offers much higher beam currents than a gallium FIB. It is used for large cross-sections, big-volume 3D tomography and samples where gallium implantation would be a problem.

Does FIB milling damage the sample?

Yes. Ions amorphise the surface, implant into it and can redeposit sputtered material. Damage is reduced by protective caps, lower currents and a final low-kV or argon polishing step.