MCM 2019
14th Multinational Congress on Microscopy

September 15-20, 2019 in Belgrade, Serbia

Days
Hours
Minutes
Seconds

Scanning Tunneling Microscope: How STM Images Individual Atoms

· 7 min read

A scanning tunneling microscope (STM) images conductive surfaces by measuring a quantum tunneling current between a sharp metal tip and the sample, with enough precision to resolve single atoms. This guide explains how the current is generated, the two main imaging modes, how tips are prepared and why STM needs vacuum and vibration isolation.

Scanning tunneling microscope definition: a microscope without lenses

A scanning tunneling microscope (STM) is an instrument that images a conductive surface by scanning an extremely sharp metal tip a fraction of a nanometre above it and measuring the tiny electrical current that flows across the gap. There are no lenses, no light and no electron beam. The image is built point by point from that current, which is so sensitive to distance that the microscope can resolve individual atoms.

The STM was developed by Gerd Binnig and Heinrich Rohrer at IBM's Zurich research laboratory in 1981, and the two shared the Nobel Prize in Physics in 1986. It was the first member of the family now called scanning probe microscopes, which also includes the atomic force microscope. Its typical lateral resolution is better than 0.1 nm and its vertical resolution reaches around 0.01 nm.

How a scanning tunneling microscope works: quantum tunneling current

In classical physics, electrons cannot cross an insulating gap such as vacuum unless they have enough energy to jump over the barrier. Quantum mechanics says otherwise. An electron's wavefunction does not stop abruptly at a surface; it decays exponentially into the gap. If a second conductor is brought close enough, there is a finite probability that electrons will tunnel through the barrier.

In an STM, the tip and sample are held roughly 0.4 to 0.7 nm apart and a small bias voltage, usually a fraction of a volt, is applied between them. The resulting tunneling current is typically in the picoampere to nanoampere range.

Why the exponential matters

The current falls by about an order of magnitude for every 0.1 nm increase in separation. That steep dependence is the secret of the STM's resolution. Almost all the current flows through the single atom at the very end of the tip that sits closest to the surface, so the effective probe is one atom wide even if the tip looks blunt under an optical microscope.

Key point: an STM does not "see" atoms as solid balls. It maps where electrons are available for tunneling at a given energy, which is closely related to, but not identical with, the positions of atomic nuclei.

Constant-current vs constant-height mode

Piezoelectric ceramics move the tip in x, y and z with sub-angstrom precision. As the tip rasters across the sample, the instrument can work in two main ways.

FeatureConstant-current modeConstant-height mode
What is held fixedTunneling current, via a feedback loopTip height above the average surface plane
What forms the imageVertical piezo movement needed to keep the current steadyVariations in the tunneling current
Scan speedSlower, limited by feedback responseFaster, useful for dynamic processes
Surface suitabilityRough or stepped surfacesVery flat, small areas only
Risk of tip crashLowHigh if the surface has steps or debris
Typical useMost routine topographic imagingFast atomic-resolution scans on flat terraces

In constant-current mode, the resulting image is often described as a topograph, but strictly it is a map of constant electron density of states. On a surface with different elements or adsorbed molecules, a bright spot can mean higher electronic density rather than a taller feature.

Scanning tunneling microscope images of atoms

The most famous scanning tunneling microscope images of atoms came early. In 1983 the Zurich group published a real-space image of the 7×7 reconstruction of the silicon (111) surface, settling a long-running debate about its structure. Since then, STM has imaged graphite and graphene lattices, metal surfaces, semiconductor dopants, molecules lying on surfaces and defects in two-dimensional materials.

Moving individual atoms

The STM can do more than look. By lowering the tip and adjusting the current, researchers can drag or lift single atoms and molecules. In 1989 IBM researchers positioned 35 xenon atoms on a nickel surface to spell the company's initials, and later work built "quantum corrals" of iron atoms on copper that revealed standing electron waves inside the ring.

Spectroscopy as well as images

In scanning tunneling spectroscopy (STS), the tip is held at one spot and the bias voltage is swept while the current is recorded. The derivative dI/dV is related to the local density of electronic states, which lets researchers measure band gaps, superconducting gaps and molecular orbitals with atomic-scale spatial resolution.

Scanning tunneling microscope tip: materials and preparation

The scanning tunneling microscope tip is the most important and most temperamental part of the instrument. A good tip ends in a single, stable atom; a poor one gives doubled features, streaks or sudden jumps in the image.

  • Tungsten is the common choice for ultra-high-vacuum work because it is stiff and can be sharpened by electrochemical etching.
  • Platinum–iridium resists oxidation, so it is popular for STMs operating in air; tips are often simply cut or sheared from wire.
  • Gold and other metals are used for specialised experiments, including some electrochemical and optical setups.

A typical preparation sequence for a tungsten tip looks like this:

  1. Etch a thin tungsten wire electrochemically in an alkaline solution until the immersed part drops off, leaving a sharp apex.
  2. Rinse the tip to remove electrolyte residue and inspect it under an optical microscope.
  3. Transfer it into the vacuum chamber and remove the surface oxide, for example by heating or ion sputtering.
  4. Condition the tip in situ with short voltage pulses or a gentle controlled contact with a clean metal surface.
  5. Test it on a well-known reference surface and repeat conditioning until atomic resolution is stable.

Experienced users accept that tips change during a session. A picked-up atom or molecule can suddenly sharpen or ruin the image, which is why reference surfaces are kept on hand.

Vacuum, vibration and temperature requirements

Because a change of 0.1 nm in gap size changes the current tenfold, any vibration of that size wrecks the image. Footsteps, building ventilation and even sound can be enough. STM designs therefore combine several layers of isolation: soft spring suspension, eddy-current damping, heavy tables and, in many labs, a separate foundation or quiet basement room.

Many research STMs operate in ultra-high vacuum so the surface stays clean for hours or days. Low-temperature instruments cooled with liquid helium to around 4 K reduce thermal drift and freeze atoms in place, which is essential for atom manipulation and high-resolution spectroscopy. At the other end of the range, STMs also work in air, in liquids and in electrochemical cells, accepting lower stability in exchange for more realistic conditions.

STM vs AFM and other microscopes

The main limitation of STM is that the sample must conduct electricity, at least weakly, because the measurement depends on a current. Insulators such as most polymers, biological tissue and thick oxides cannot be imaged directly. The atomic force microscope was developed partly to solve this by measuring forces instead of current.

PropertySTMAFMSEM
SignalTunneling currentTip–sample forceElectrons emitted from the sample
Sample requirementConductive or semiconductingAlmost any solid surfaceUsually conductive or coated
Best resolutionAtomic, routinelyAtomic in special conditionsNanometre scale
InformationSurface electronic structureTopography, mechanicsMorphology, composition with EDS
Field of viewNanometres to a few micrometresUp to about 100 micrometresMillimetres down to nanometres

For a broader overview of where scanning probes sit among optical and electron techniques, see our guide to types of microscopes.

Scanning tunneling microscope cost and who uses one

The scanning tunneling microscope cost varies enormously. Compact benchtop instruments that work in air are made for teaching and can show atomic lattices on graphite at a price within reach of many university departments. Research systems with ultra-high vacuum, cryogenic cooling, sample preparation chambers and vibration-isolated rooms are a major capital investment and require trained staff to run. Prices depend heavily on configuration, so institutions request quotes from manufacturers rather than relying on list figures.

Typical users include surface physicists, materials scientists studying two-dimensional materials and catalysts, chemists working on single molecules, and research groups exploring quantum devices.

Conclusion

The scanning tunneling microscope turns a quantum effect into a practical tool: a sharp tip, a tiny voltage and a current that changes tenfold with each tenth of a nanometre. That sensitivity lets STM image and even rearrange individual atoms on conductive surfaces, at the cost of demanding tips, vacuum and vibration control. For non-conductive samples, AFM is usually the better choice, but for electronic structure at the atomic scale, STM remains the reference technique.

Frequently asked questions

How does a scanning tunneling microscope work?

A sharp metal tip is brought within about half a nanometre of a conductive surface and a small voltage is applied. Electrons tunnel across the gap, and the current, which depends exponentially on distance, is recorded while the tip scans to build an image.

Who invented the scanning tunneling microscope?

Gerd Binnig and Heinrich Rohrer developed it at IBM Zurich in 1981. They received the Nobel Prize in Physics for the invention in 1986.

Can a scanning tunneling microscope image insulators?

Not directly, because STM needs a tunneling current to flow between tip and sample. Very thin insulating films on a metal can sometimes be imaged, but bulk insulators require AFM or another technique.

What is an STM tip made of?

Most tips are tungsten, sharpened by electrochemical etching, or platinum-iridium, which is cut from wire and resists oxidation in air. The working end is ideally a single atom.

What is the difference between STM and AFM?

STM measures a tunneling current and requires conductive samples, while AFM measures forces on a cantilever tip and works on almost any surface. STM gives atomic resolution more routinely and also probes electronic structure.

How much does a scanning tunneling microscope cost?

It depends on the configuration. Benchtop teaching instruments are relatively affordable, while ultra-high-vacuum, low-temperature research systems are a major investment priced individually by manufacturers.